5. Specifications
5.1 BioHarness Module
Parameter
Specification
Heart Rate:
25 240 Beats per Minute ± 1 BPM
Breathing Rate:
3 70 Breaths per Minute
Skin Temperature:
10 60 °C ± 0.2°C in range 30 40°C
Battery Duration:
24 hrs Standby Mode, 18 hrs Active Mode
Charge Time:
3 hours to 100%, 1 hour to 90%
Battery Life:
Capacity 80% after 300 deep discharge cycles (4500hrs of use)
Storage Life:
6 months between charges (for optimal recharge monthly)
Strap Sizes:
XS (22 26”), S (26 32.5”), M (32.5 – 39”), L (39 45.5”), XL (45.5 –
52”)
Wash durability:
Typically 50 washes with BioHarness Module removed
Environmental Operating:
Humidity:
5 90 %RH
Temperature:
-10 50°C
Environmental Storage:
Humidity:
5 90 %RH
Temperature (< 1 month):
(Power off) -20°C 45°C
Temperature (< 6 months):
(Power off) -20°C 35°C
Water Resistance (Device):
IPX7 (immersion for up to 30minutes in 1 m of water)
Bluetooth Compliance:
Version 2.0 + EDR
Operating Frequency:
2.4 2.835 GHz
Output Power:
2 mW